HyPrep Plus (Model 4052-054P) Specifications
Throughput: Tube-to-Plate Dilution; 96 wells, less than 24-minutes. (Varies with application). (Overall throughput may be increased by system's multitasking capability.) Delivery Range: 1µl to 1,000µl (in increments of 1µl). Precision: Coefficient of variation (CV) ±5.0%. Accuracy: 1µl to 9µl: <=2%; 10µl to 1000µl: <=1%. Diluter Data: 2 High-precision, positive displacement cylinders with solenoid valve; range: 0-50µl and 0-1000µl Interface: RS-232 Serial Port. Power Requirements: 115/230V, ±10%, 60/50 Hz Dimensions: Probe Work-area: 13.63" (34.54cm) L x 5.80" (14.70cm) W; HyPrep Plus Cabinet: 22.75" (57.8cm) x 17.25" (43.8cm) D x 13.25" (33.7cm) H; Weight: 40Lbs (18.0 Kg). |
|
Computer Requirements: 386 or 486 with 2 Meg of RAM and 3.5" Floppy Disk Drive; 486 required for multitasking; (System's MULTITEST PLUS Software included; computer not included) Monitor: 14" VGA Monitor (color/monochrome); not included. Connections: 1 Serial Port required. (1 Parallel and 2 serial ports required when a printer, mouse or bar code reader are desired.) Operating System: Min. MS-DOS® 5.0 or higher; Windows 95 for multitasking.
Process Temperature
Range: Ambient to 37°C and above; User
programmable; Factory default set at 37° C. Plate Temperature Uniformity: ±0.5°C from set temperature. Controller Type: Proportional Integral Derivative (PID).
|
Capacity: 1 standard 96-well microplate. Power Requirements: 120V/60Hz or 220V/50Hz, non-selectable (must specify when ordered.) ThermoPrep 2 Option Only: Shaking Speed: Up to 2500 RPMs (User programmable); Orbital shaking pattern.
Sample Size: 1 standard 96 well microplate. Applications: Absorbance, Concentration Curve, Single-point Concentration and Qualitative (Cutoff) Tests. Photometric Method: Monochromatic & Bichromatic; wavelength range: 400nm to 700nm; Measurement range: 0 to 3.0 Optical Density (OD); Indication up to 4.0 OD. Reading Time; Full Plate: Less than 20 seconds, bichromatic. Filter Capacity: 6-position filter wheel, equipped with six standard filters: 405nm, 450nm, 490nm, 540nm, 630nm and 650nm.
|
Accuracy/Linearity: ±2.0% or ±0.010 OD whichever is greater, in the range of 0.1 to 2.5 OD at 405nm / ±1.25% of range, in the 0 to 2.5 OD range at 405nm. Reproducibility: ±1.25% of range, in the 0 to 2.5 OD range at 405nm. Light Source: Tungsten Halogen Lamp; Computer Built-in: 486 with 2 MB RAM (or equivalent); 3.5 Floppy Disk Drive. Display: Half page LCD Display; Displays alphanumeric and graphic data. Keyboard: IBM® Style, 101 key compatible, with 12 color-coded user interface keys. Interface Ports: 1 Parallel; 2 Serial Ports; 1 Bar Code Reader Interface Port; Bidirectional RS-232 Interface Port; 1 External VGA Monitor Port. Power Supply: 115V/230V ±10%; 60/50Hz. IBM is a registered trademark of International Business Machines |